I have a Gow-Mac GC chromatograph. How should I connect it with Clarity station?

For Gow-Mac GC chromatograph use asymmetrical connection as shown on the picture below.

Connect Brown wire (Pink tubular end sleeve) to Terminal (+) and White wire (Grey tubular end sleeve) together with Yellow wire (shielding) to Terminal (-).

Do not connect anything to Terminal marked D on picture, it may cause the damage of the A/D converter.

Set also the Signal range at Method Setup Acquisition tab to the lowest possible value.

gow-mac-wiring

Send comment15.1.2019
 