I have configured FLD so that two Signals are on Instrument 1 and other two on Instrument2. When I open Instrument 2 and go to Device Monitor it does not display any signal.

It is not posible even technically for the device to run on two independent timebases.
This problem applies generally to all Control modules. Some devices contain parts "subdevices", such as Signals in FLD that cannot cannot be used on two independent chromatography sets at once. It is usually not technically possible. The most recent Control modules will show a warning during the configuration procedure when you attempt to split indivisible device between multiple Clarity Instruments.