How to set proper Sampling Rate?

The Sample Rate needs to be set to provide enough points to adequately describe the peaks of interest. Recomended is at least 30 points per the narrowest peak width. The default value for DataApex A/D converters is 10 Hz. This is suitable for most HPLC and packed column GC applications, the capillary GC and CE applications may need higher sample rates.

Accordingly to Sampling Rate, the Global Peak Width parameter has to be also amended. See also FAQ How to set proper Peak Width parameter?.