INT7 - PCI Board of A/D Integration Converters

INT7 - internal PCI A/D Converter

INT7 is an internal 24-bit A/D converter board that measures the voltages of chromatographic and electrochemical detectors. It contains up to 4 data-acquisition channels on one board (one and two channel versions are also available). The channels are completely independent, sigma delta integrating A/D converters, used for processing positive and negative voltages. The board includes four digital inputs and eight digital outputs. The first four digital outputs are also equipped also with relay contacts. The board is manufactured using the SMT montage and provides excellent reliability and temperature stability.

Full size PCI 2.0, 32 bit PCI slot is required. Low profile or PCI Express slots cannot be used.

INT7 is the direct successor of our classic INT5 ISA card. The two share an identical analog channel design and thus, identical basic electric parameters.


Advantages of the INT7 A/D converter:

  • No loss of input signal during integration. There are no time delays during which the converter fails to integrate the input signal.
  • There is no continuous servicing of the input analog switches, which would otherwise cause errors (offset, noise, etc.).
  • The amount of data entering the computer is substantially smaller, since individual data items already represent partial integrals.

Next steps

Technical data:

INT7
Converter type 24 bit integrating, delta-sigma integration
Analog inputs differential
Number of channels 1, 2 or 4 fully independent, mutually isolated
Input range unipolar and bipolar, each channel independentely: 156, 1250 or 12560 mV
Input impedance > 10 MOhm
Sample rate 10, 25, 50, 100 Hz
Non-linearity < 0.0015 %
Temperature dependence < 10 ppm/°C
Digital inputs 4 TTL (5 V), the first two can be controlled using the built-in opto-couplers
Digital outputs 8 TTL (5V), maximum current load 20 mA), the first four control relays
INT7

Effective resolution for various acquisition speeds and input ranges:

10 V 1 250 mV 156 mV
10 Hz 20 bit 19 bit 16 bit
50 Hz 18 bit 17 bit 15 bit
100 Hz 17 bit 15 bit 14 bit
10 V 1 250 mV 156 mV

Possible combinations:

WIN XP WIN 2000 WIN NT Win 98, Me
INT7 Clarity
Clarity Lite
CSW32
Clarity
Clarity Lite
CSW32
Clarity
CSW32
Clarity (until 2.4.1)
CSW32
WIN XP WIN 2000 WIN NT Win 95, 98, Me